PiCCO Trouble Shooting Guide | For PulsioFlex with PiCCO module

Arterial blood pressure signal

Pressure curve error [4/5]

not displayed, flat, noisy or no dicrotic notch detection

1

Replace pressure transfer cable 

2

Perform pressure zeroing